    MicroBio Cross Contamination Risk Statement
    Technical Resources
    MicroBio Cross Contamination Risk Statement
    Published

    MicroBio Cross Contamination Risk Statement

    Stephen Plumridge By Stephen Plumridge

    The MicroBio range of bioaerosol samplers are designed to offer minimal chances of cross

    contamination within all applications, so long as operational precautions are taken.

    Definition

    Cross-contamination is what happens when bacteria or other microorganisms are unintentionally

    transferred from one object to another. Moving any equipment between cleanrooms or between

    different cleanroom classes will always carry a risk of cross contamination.

    Avoiding Cross Contamination

    To avoid cross contamination, the air sampler should be designated for use only in one dedicated

    zone and only for one cleanroom class. Different zones or classes should each have their own air

    sampler.

    If moving the same air sampler between zones within the same cleanroom class, then cleaning

    protocols should be used when the device leaves the zone and when entering the next zone.

    Such protocols are under the control of users only, but we have some recommendations below.

    Cleaning of MicroBio Air Samplers

    The devices can be cleaned using most methods employed within cleanrooms, except the use of

    caustic based cleaning agents. IPA wipes or hydrogen peroxide vapour (HPV) based cleaning can

    be regularly used. Full guidance is in the operating manuals. For the air inlet and exhaust points,

    cleaning swabs can be used on the inlet and exhaust paths along with IPA cleaning sprays.

    Alternatively HPV cleaning can be used.

    Mechanisms for Cross Contamination

    All surfaces are easily cleaned. The air flow path on MicroBio Air Sampler passes direct from the

    blower inlet, through the blades and is directly exhausted. It does not pass over any other

    components within the structure of the air sampler.

    Limitation

    MicroBio air samplers do not incorporate HEPA filtering. HEPA filtering only needs to be

    considered if the same single air sampler is to be used across multiple cleanroom classes and

    between use thorough cleaning is not possible.

    Disclaimer

    Cantium Scientific Limited has no control of the operation of the products at customer facilities. It

    is the customer’s responsibility to determine the suitability of the MicroBio air sampler to their

    application.